6.4 Input Sensitivity and Calibration
The input sensitivity can be manually changed in the text field or it can be calibrated using the
calibration function and a sensor calibrator. Click the button to launch the Calibration
interface.
6.4.1 Calibration Interface
Step 1 Adjust the calibration parameters
The defaults values are:
Averaging time: 5 s
Frequency: 1 kHz
Calibrator Level: 94 dB
Step 2 Install the calibrator device on the microphone
Step 3 Click START
After the average time is elapsed, the sensitivity value will update.
Step 4 Click OK to accept the sensitivity value